A flue gas analyser makes safety decisions — so a reading is only worth anything if the instrument is healthy and in calibration. An analyser that's drifted can pass a dangerous appliance or fail a good one. This guide covers keeping one reliable. It's study material; only a Gas Safe registered engineer may carry out the work.

The sensors age — that's the core issue

Most analysers use electrochemical sensors for CO and O₂. These react chemically with the gas to produce a tiny electrical signal, and that chemistry wears out over time whether or not the instrument is used — typically lasting a few years. As a sensor ages it drifts, so the instrument needs checking against known gas and eventually re-sensoring.

Calibration with certified gas

Calibration means comparing the analyser against certified reference gas of a known concentration and correcting any drift. The accepted requirement is calibration at least annually (and per the manufacturer), carried out by an approved facility that issues a calibration certificate. A reading from an out-of-calibration analyser isn't valid — and in an assessment, "is it in calibration?" is a question you should ask before trusting any figure.

Zeroing is not calibration. Zeroing in fresh air on power-up sets the baseline for that session; calibration is the periodic check against certified gas. You do the first every time you use it; the second on a schedule.

Everyday care

Response and recovery

Give the analyser time to respond (reach a steady reading in the flue) and to recover (return to zero in fresh air) — rushing either gives a misleading figure. A sensor that's slow to recover may be tired and due for attention.

  1. Electrochemical sensors age and drift over a few years, used or not.
  2. Calibrate at least annually against certified gas; keep the certificate.
  3. A reading from an out-of-calibration analyser is not valid.
  4. Zeroing ≠ calibration: zero every use, calibrate on schedule.
  5. Fresh-air purge before switching off to protect the sensors.
  6. Maintain water trap, filter, probe and hose; check for leaks.
  7. Allow response and recovery time for honest readings.

10-Question Mock Test

Click an option to see whether you got it right. Explanations appear instantly — no submitting at the end.

Your score: 0 / 10
Question 1 of 10
CO and O₂ in most analysers are measured by:
Question 2 of 10
Do analyser sensors age even if the instrument is rarely used?
Question 3 of 10
How often must an analyser typically be calibrated?
Question 4 of 10
Calibration is carried out against:
Question 5 of 10
Zeroing in fresh air on start-up is the same as calibration. True or false?
Question 6 of 10
A reading taken with an out-of-calibration analyser is:
Question 7 of 10
Running the analyser in fresh air before switching off:
Question 8 of 10
The particulate filter should be:
Question 9 of 10
A split probe hose or perished seal could cause:
Question 10 of 10
A sensor that is slow to return to zero in fresh air may indicate:

An analyser you can trust is a pass you can defend.

PlumbMate drills the CPA1 care-and-calibration detail with quizzes and spaced repetition mapped to the gas ACS tickets.

🔒 PlumbMate Gas — coming soon

Full ACS revision — CCN1, CPA1, CKR1, HTR1 & CENWAT · £29.99/year · Launching soon

Prefer to browse first? Back to the Gas Blog →